IORT safety and dosimetry
Check whether IORT commissioning, dosimetry, shielding and delivery safety are represented; do not treat unrelated external-beam simulation as equivalent.
SBRT terminology and planning context
Check SBRT terminology, uncertainty, image guidance and QA context; teaching images must not be presented as verified clinical plans.
Reference dosimetry boundaries
Keep illustrative dose graphics distinct from reference dosimetry and absorbed-dose-to-water calibration.
Corrigendum applicability
Use the corrected Rev. 1 edition before implementing or quoting numerical reference-dosimetry data.
MRI geometry and registration
State MRI geometric distortion and registration limits; do not imply registration validation without measured or dataset-based tests.
Planning-system QA distinction
Keep educational plan visualisation clearly distinct from a commissioned treatment planning system and from TPS acceptance/commissioning claims.
Physics training traceability
Map teaching content to named curriculum domains without claiming complete curriculum coverage.
Independent MU/dose verification
Do not imply independent dose/MU verification unless an independently tested calculation pathway exists.
CBCT image-quality and localisation context
Distinguish CBCT localisation/image-quality QA from diagnostic imaging and from contouring accuracy.
Radiation oncologist curriculum mapping
Map educational pathways to published radiation-oncology competencies and identify gaps explicitly.
Afterloader safety and applicator QA
Separate applicator/source geometry teaching from clinical afterloader QA, source checks and delivery safety.
Trial protocol QA traceability
Clinical-trial examples must link to protocol-specific requirements and must not infer eligibility or credentialing from generic teaching content.
Measurement-based patient-specific QA
Dose rendering must not be presented as measurement-based patient-specific IMRT QA unless measured-comparison functionality exists and is validated.
Standard ROI and plan metric names
Use versioned standard structure/metric nomenclature; do not silently convert unsupported synonyms.
Structure labels in brachytherapy
Apply TG-263 naming only where the relevant brachytherapy structure or concept is covered; preserve unsupported names explicitly.
ROI display colour provenance
Treat DICOM ROI display colour as optional display metadata, not as a mandatory anatomical colour standard.
Stereotactic prescription and reporting
Distinguish prescribed, planned/simulated and measured quantities in stereotactic treatment reporting.
Small-photon-beam modelling claims
Educational small-field graphics must not be presented as calibrated small-field dosimetry or validated beam modelling.
Registration and fusion uncertainty
State registration frame, method and uncertainty; validate registration independently before using transformed geometry as authoritative.
Small-field reference dosimetry
Keep small-field reference/output-factor dosimetry distinct from illustrative beam geometry and use appropriate detectors/methods for quantitative claims.
Cervical brachy reporting
Use cervix brachytherapy prescription/reporting terminology in the context defined by ICRU 89; do not generalise it to unrelated sites.
Risk-based quality management
Use process mapping and risk analysis for quality management; do not mislabel TG-100 as a software-only QA standard.
Radiation therapist education
Map RTT learning content to the ESTRO curriculum without claiming full curricular coverage.
IMRT reporting nomenclature
Use ICRU 83 terminology when presenting IMRT prescription/reporting concepts; clearly distinguish planned and measured quantities.
Historical margin nomenclature
Use ICRU 62 as historical margin/reporting terminology and distinguish it from modern automated margin operations.
Foundational target volume terminology
Use ICRU 50 for foundational target-volume/reporting concepts, not as an algorithmic contouring prescription.
DICOM geometry and identity
Test patient image spacing, orientation, position, frame identity and structure-set references against the exact DICOM PS3.3 modules.
Brachy RT DICOM interoperability
Test brachy source/channel/application-setup and contour references against the exact DICOM RT modules used by the implementation.
Image provenance and coordinate mapping
Use DICOM patient-coordinate definitions as the boundary for patient image geometry; constructed atlas coordinates must not be presented as patient DICOM geometry.